ISO5 GMP Containment Powder Weighting Booth
Product Details:
Place of Origin: CHINA
Brand Name: HaoAir
Certification: CE,
Payment & Shipping Terms:
Minimum Order Quantity: 1SET
Price: NEGOATION
Packaging Details: PLOWOOD PACKING
Delivery Time: 8 WORKING DAYS
Payment Terms: Ex works/FOB/CIF/ CNF
Supply Ability: 200Sets/month
- Description
- Inquiry
Weighting Booth Introduction
Weighting Booth also know as laminar flow booth, dispensing booth is for the pharmaceutical, microbiological research and scientific experiments and other places dedicated local purification equipment, which provides a vertical unidirectional airflow, some clean air circulation in the work area, partly discharged to the nearby area, so that the work area produce negative pressure to prevent cross contamination in order to ensure the high clean class level of the working area.
Using Environment and Conditions:
Temperature: 5 ℃ -40 ℃
Relative humidity: less than 85%
Atmospheric pressure: 86kPa-106kPa
Power supply: AC220V, single phase, 50Hz.
Standard Model Technology Parameter:
Weighting Booth Parameter :
Model | HA-DB2525 |
External size(mm) | 2400x2800x2650mm |
Internal size(mm) | 2500X2500X2150mm |
Cleanliness: | Class100 |
HEPA Filter | 99.995@0.3um |
Pre-filter | 75%@5um |
Air velocity | 0.2-0.5m/s |
Noise | 57-65dB |
Material | Stainless steel 201/304 or steel with power coated |
Power | 380V/50hz |
Standard | (FS 209E)USA . |
Customer’s design is available.
Factory test:
All of equipment is individually factory tested for safety and performance in accordance with international Standards. Each unit is shipped with a documentation outlining the tests undertaken and the units individual results for each unit.
Factory test include:
— functional test and visual inspection
— electrical safety tests
— air velocity testing
— purification rank test (if need)
Warranty:
Our equipment is warranty for 1 year excluding consumable parts and accessories.
All equipment is shipped with a comprehensive use’s manual complete with a report documenting all test procedures.
Additional IO/OQ/GMP document is available upon request.
Contact our sales representative for specific warranty details or document request.
Consumable parts illustration:
1: Pre-filter: each one should be replacement in each 6 month, but it can refresh no more three times.
2: HEPA filter: each one should be replacement in each half and one years.